step_sec_mals() creates a specification of a recipe step that processes multi-angle light scattering (MALS) detector signals to determine absolute molecular weight and radius of gyration at each elution slice.

Usage

step_sec_mals(
  recipe,
  mals_col = NULL,
  concentration_col = NULL,
  dn_dc = NULL,
  dn_dc_column = NULL,
  wavelength = 658,
  solvent_ri = 1.333,
  angles = 90,
  formalism = c("zimm", "debye", "berry"),
  calibration_constant = NULL,
  output_mw = "mw_mals",
  output_rg = "rg",
  min_signal = 0.01,
  role = NA,
  trained = FALSE,
  skip = FALSE,
  id = recipes::rand_id("sec_mals")
)

Arguments

recipe

A recipe object.

mals_col

Name of the MALS detector measure column. For single-angle detectors (RALS/LALS), this is the only signal needed.

concentration_col

Name of the concentration measure column (from step_sec_concentration() or similar).

dn_dc

Refractive index increment (mL/g). Required for absolute MW.

dn_dc_column

Column containing sample-specific dn/dc values.

wavelength

Laser wavelength in nm. Default is 658 (common for MALS).

solvent_ri

Solvent refractive index. Default is 1.333 (water). Common values: water = 1.333, THF = 1.407, toluene = 1.497.

angles

Numeric vector of detection angles in degrees. For single-angle detectors, provide just one value (e.g., 90 for RALS). Default assumes a 90-degree detector.

formalism

Angular extrapolation method for multi-angle data: "zimm" (default), "debye", or "berry".

calibration_constant

MALS instrument calibration constant. If NULL, results are in relative units. Obtain from toluene standard calibration.

output_mw

Name for the molecular weight output column. Default is "mw_mals".

output_rg

Name for the radius of gyration output column. Default is "rg". Only calculated if multiple angles are provided.

min_signal

Minimum signal threshold (as fraction of max) below which MW is set to NA. Default is 0.01.

role

Role for generated columns.

trained

Logical indicating if the step has been trained.

skip

Logical. Should the step be skipped when baking?

id

Unique step identifier.

Value

An updated recipe with the new step added, containing absolute MW (and optionally Rg) at each elution point.

Details

Light scattering provides absolute molecular weight without calibration standards. The fundamental relationship is:

$$\frac{K \cdot c}{R(\theta)} = \frac{1}{M_w} \cdot P(\theta) + 2A_2 c$$

where:

  • K is the optical constant

  • c is the concentration

  • R(theta) is the excess Rayleigh ratio

  • Mw is the weight-average molecular weight

  • P(theta) is the particle scattering function

  • A2 is the second virial coefficient

The optical constant K is calculated as: $$K = \frac{4\pi^2 n_0^2 (dn/dc)^2}{N_A \lambda^4}$$

Formalisms for Angular Extrapolation:

  • Zimm: Kc/R vs sin^2(theta/2) - best for random coils

  • Debye: Kc/R vs sin^2(theta/2) - similar to Zimm

  • Berry: sqrt(Kc/R) vs sin^2(theta/2) - better for large particles

Single-Angle vs Multi-Angle:

  • Single angle (RALS/LALS): Provides Mw only, assumes P(theta) ~ 1

  • Multi-angle (MALS): Provides both Mw and Rg from angular dependence

Note

Accurate results require:

  • Known and accurate dn/dc value

  • Calibrated instrument (calibration_constant from toluene)

  • Accurate concentration from RI detector

  • Clean baseline and aligned detectors

Examples

if (FALSE) { # \dontrun{
library(recipes)
library(measure)

# Single-angle (90 degree) light scattering
rec <- recipe(~., data = sec_triple_detect) |>
  step_measure_input_long(ri_signal, location = vars(elution_time), col_name = "ri") |>
  step_measure_input_long(mals_signal, location = vars(elution_time), col_name = "mals") |>
  step_sec_baseline() |>
  step_sec_ri(dn_dc = 0.185) |>
  step_sec_concentration(detector = "ri", injection_mass = 0.2, flow_rate = 1.0) |>
  step_sec_mals(
    mals_col = "mals",
    concentration_col = "ri",
    dn_dc = 0.185,
    wavelength = 658,
    angles = 90
  ) |>
  prep()
} # }